双乳被老汉揉搓a毛片免费观看|国产精品9999久久久久|少妇厨房愉情理伦bd在线观看|国产精品无码翘臀在线观看|亚洲日韩精品一区二区三区

 歡迎訪問(wèn)濟(jì)南科爾超聲波設(shè)備有限公司網(wǎng)站!

專注生產(chǎn)工業(yè)用超聲波清洗機(jī)

讓工業(yè)清洗無(wú)難事

電話

業(yè)務(wù)咨詢電話:

18663767799

你所在位置: 首頁(yè)-新聞動(dòng)態(tài)-新聞資訊多槽手動(dòng)超聲波清洗機(jī)的幾種部件維修方法
新聞資訊

多槽手動(dòng)超聲波清洗機(jī)的幾種部件維修方法

作者:admin來(lái)源:http://www.glwh.com.cn/時(shí)間:2023-02-02

如果使用不當(dāng)或不注意維護(hù)保養(yǎng)工作的話,多槽手動(dòng)超聲波清洗機(jī)是很容易出現(xiàn)故障損壞的,常見(jiàn)故障有換能器損壞、功率管損壞、電源無(wú)反應(yīng)、穩(wěn)壓管損壞、沒(méi)有超聲波輸出等,需要排查出故障原因進(jìn)行維修或更換。下面來(lái)詳細(xì)講解下。
If it is not used properly or the maintenance work is not paid attention to, the multi-slot manual ultrasonic cleaning machine is prone to failure and damage. Common failures include damage to the transducer, damage to the power tube, lack of response to the power supply, damage to the stabilizer tube, and no ultrasonic output. It is necessary to find out the cause of the failure for repair or replacement. Let's explain it in detail.
換能器損壞
The transducer is damaged
故障原因:多槽手動(dòng)超聲波清洗機(jī)長(zhǎng)時(shí)間處于開(kāi)機(jī)使用狀態(tài),溫度逐漸上升,導(dǎo)致膠體融化換能器脫落或換能器陶瓷部分?jǐn)嗔选?/div>
Cause of failure: The multi-slot manual ultrasonic cleaning machine has been in use for a long time, and the temperature is gradually rising, which causes the gel melting transducer to fall off or the ceramic part of the transducer to break.
維修方法:用搖表測(cè)量換能器的絕緣強(qiáng)度,絕緣強(qiáng)度在200MΩ以下已無(wú)法使用,更換新的換能器即可。
Maintenance method: use a megger to measure the insulation strength of the transducer. If the insulation strength is below 200M Ω, it can no longer be used. Replace the transducer with a new one.
功率管損壞
Power tube damaged
故障原因:多槽手動(dòng)超聲波清洗機(jī)長(zhǎng)期不間斷使用、清洗液體太少都會(huì)導(dǎo)致功率管短路。
Cause of failure: The power pipe will be short-circuited if the multi-slot manual ultrasonic cleaner is used continuously for a long time and the cleaning liquid is too little.
維修方法:功率管在主板上連接時(shí),用萬(wàn)用表測(cè)量功率管兩側(cè)管腳的阻值,正常情況下應(yīng)在22Ω左右,否則需要更換功率管。
Maintenance method: When the power tube is connected on the main board, use a multimeter to measure the resistance value of the pins on both sides of the power tube, which should be around 22 Ω under normal conditions, otherwise the power tube needs to be replaced.
開(kāi)機(jī)后電源無(wú)反應(yīng),不顯示
The power supply does not respond and does not display after startup
故障原因:接地線與火線或零線混用導(dǎo)致電源座內(nèi)保險(xiǎn)損壞,或者多槽手動(dòng)超聲波清洗機(jī)元器件老化導(dǎo)致短路。
Fault cause: The fuse in the power supply socket is damaged due to the mixed use of the grounding wire and the live wire or the zero wire, or the aging of the components of the multi-slot manual ultrasonic cleaner causes a short circuit.
維修方法:取出保險(xiǎn)看有無(wú)斷裂,并用萬(wàn)用表通斷檔測(cè)量是否斷開(kāi),如果損壞了要更換新的保險(xiǎn)。
Maintenance method: take out the fuse to see if it is broken, and use a multimeter to measure whether it is broken. If it is damaged, replace it with a new one.
多槽手動(dòng)超聲波清洗機(jī)
穩(wěn)壓管或橋損壞
The stabilizer tube or bridge is damaged
故障原因:功率管阻值小或短路后開(kāi)機(jī)會(huì)造成穩(wěn)壓管和二極管整流電路損壞。
Cause of failure: The power tube resistance is small or the power supply is started after a short circuit, which will cause damage to the voltage regulator and diode rectifier circuit.
維修方法:用萬(wàn)用表二極管檔測(cè)量,正向阻值70Ω左右,反向∞。如數(shù)值偏差過(guò)大或正向阻值也是∞,更換新穩(wěn)壓管。
Maintenance method: measure with the diode gear of a multimeter, and the forward resistance is about 70 Ω, and the reverse resistance is ∞. If the value deviation is too large or the positive resistance value is also ∞, replace the regulator with a new one.
電感、隔離變壓器損壞
Damage of inductance and isolation transformer
故障原因:多槽手動(dòng)超聲波清洗機(jī)長(zhǎng)時(shí)間工作導(dǎo)致電感、變壓器等散熱元件因溫度過(guò)高而融化、燒毀。
Cause of failure: The multi-slot manual ultrasonic cleaning machine worked for a long time, resulting in the melting and burning of the inductor, transformer and other cooling elements due to high temperature.
維修方法:一般這兩個(gè)器件損壞都是被燒壞了,可以直接看出燒壞的痕跡,無(wú)法維修,只能更換新器件。
Maintenance method: Generally, these two components are damaged by burning, and the burning trace can be seen directly. They can not be repaired, and can only be replaced with new components.
沒(méi)有超聲波輸出
No ultrasonic output
故障原因:可能是換能器與超聲波功率板的連接插頭松脫、保險(xiǎn)絲熔斷、超聲功率發(fā)生器故障、換能器故障。
Fault cause: The connection plug between the transducer and the ultrasonic power board is loose, the fuse is blown, the ultrasonic power generator is faulty, and the transducer is faulty.
維修方法:逐步排查問(wèn)題的原因,然后進(jìn)行維修或更換。
Maintenance method: gradually find out the cause of the problem, and then repair or replace it.
任何東西都有其使用的壽命,多槽手動(dòng)超聲波清洗機(jī)如果使用不當(dāng)很容易出現(xiàn)故障,即使維修好了其使用壽命也會(huì)大打折扣,因此我們要注意多槽手動(dòng)超聲波清洗機(jī)的維護(hù)保養(yǎng)工作。更多相關(guān)事項(xiàng)就來(lái)我們網(wǎng)站http://www.glwh.com.cn咨詢吧!
Everything has its service life. If the multi-slot manual ultrasonic cleaner is not used properly, it is easy to fail. Even if it is repaired, its service life will be greatly reduced. Therefore, we should pay attention to the maintenance of the multi-slot manual ultrasonic cleaner. Come to our website for more relevant matters http://www.glwh.com.cn Ask!
主站蜘蛛池模板: 精品国偷自产在线视频99| 18禁美女裸身无遮挡免费网站| 国产人妻鲁鲁一区二区| 无码熟妇人妻av| 一本加勒比hezyo无码人妻| 尤物在线精品视频| 精品国产成人a区在线观看| 无码激情亚洲一区| 无码国产精品一区二区vr| 久久成人国产精品免费| 国产午夜精品一区二区三区老 | 精品国产免费观看久久久| 一本久道综合在线无码88| 香蕉久久夜色精品国产| 米奇777超碰欧美日韩亚洲| 久热这里只精品99国产6| 波多野结衣一区二区免费视频| 无线日本视频精品| 国产熟妇乱xxxxx大屁股网| 国产精品99久久久久久董美香| 18禁美女裸体免费网站| 久久久噜噜噜久久免费| 国产熟妇久久777777| 国产精品午夜剧场免费观看| 亚洲色爱免费观看视频| 成人免费视频视频在线观看 免费| 国产精品人成视频免费播放| 国精产品一区一区三区免费视频| 男人扒开女人双腿猛进视频| 国产愉拍精品手机| 奇米精品视频一区二区三区| 成人午夜精品无码区| 五月丁香啪啪| 伊人久久大香线蕉av网站| 久久狼人大香伊蕉国产| 日日噜噜夜夜狠狠久久丁香五月 | 久久人妻无码一区二区| 四虎影视无码永久免费| 亚洲中文字幕不卡无码| 男人吃奶摸下挵进去好爽| 大帝av在线一区二区三区|